صفحه 1:
صفحه 2:
بررسي دقت در روشهاي کمي
صفحه 3:
° Accuracy
° Precision
° Reportable range
Be Xess acs Cocm Gn Key ai
° Analytical sensitivity
* Analytical specificity
° Stability
performance لاعطاه اا
characteristic required for _ test
performance
صفحه 4:
*Imprecision - Dispersion 01
independent Packt Lay 3
measurements obtained 111011
specified conditions.
*Precision - Closeness of agreement
between independent
test/measurement results obtained
under stipulated conditions.
صفحه 5:
Repeatability conditions : Conditions
where independent test results are
obtained with the same method ده
identical test material in the same
laboratory by the same operator using the
same equipment within a short interval of
time (ISO 3534-1); NOTE: Formerly termed
Visa eae (ie
صفحه 6:
Intermediate precision conditions : Where
test results or measurement results are
obtained with the same method, on identical
test/measurement items in the same test
facility, under some different operating
condition; NOTES: a) There are _ four
elements to the operating conditions: ‘ime,
calibration, operator, and equipment; b) the
changed elements in operating conditions
must be noted; this could include precision
estimates commonly called, for example
“between-run,” “within-day,” “between-day,”
“within-device,” and “within-laboratory.”
صفحه 7:
Reproducibility conditions -
Conditions where test results are
obtained with the same method on
identical test items in different
۱۳:1۰ ۹:۱۵ ات كك لانن نت
operators using لانتو تالت
equipment (ISO 5725-1)
صفحه 8:
Run - An interval within which the trueness
and precision of a testing system is
expected to be stable, but cannot be
greater than 24 hours or less than the
frequency recommended 7 ۹3
manufacturer (US CFR493 February 28,
1992);° NOTE: (ISO3534-1/93-2.48) defines
“run” as follows: In a series of observations
of a qualitative characteristic, the
اأمتحتدءاصتست سد 01 عمدع تستتصعه Me Cel
the same attribute is called a “run.”
صفحه 9:
نمونه هاى an! استقاده
*محلولهاى استاندارد ماتريكس آنها نسبت به نمونه انسانى ساده تر
eed ل ا ل 0
og ل ل ا ا ا اا
سا از ule |) 4s A Ll
همچون پایدارکنندهها. لیوفیلیزاسیون ۰ برخی افزودنیها برای
7
oreasy ای SEES Gs dogs Galo
کرد مثل آزمایشات دوبل در یک روز
صفحه 10:
"تعداد و غلظت نمونه ها بر اساس غلظت تصمیم گیری بالینی
*بعنوان مثال براى كلوكز مىتوان 3 نمونه در غلظتهاى
ا ۱
تشخیصی دیابت) و 160mg/dL 6 ل 0
21 انتخاب نمود و برای کلسترول 2 نمونه در مقادیر
ا ۱
صفحه 11:
0روز- هر روز2 سري کاري- در هر سري 2 نمونه بصورت دوبل
*بررسی دقت دستگاهها
گروز هر روز 2 نمونه بصورت چهارتايي آزمایش ميشود.
*بررسی دقت کیتها توسط واردکنندگان
که قبلا توليد و توزیع شده اند
صفحه 12:
CLSI EP5-A2
Evaluation of Precision
Performance of Quantitative
Measurement Methods;
Approved Guideline—Second
Edition
صفحه 13:
آشنايى با روش:
07 0 ا ا I Neee Ne IES
aie” پس از تسلط بر کلیه جولنب
صفحه 14:
بر ری تب
Se ST BIS Bitte PISA Ee rene eel
EE OL Ie wl ee SIE DY YS Sree
كار حداقل بايد دوساعت باشد. در هس سرى كارى نمونه هاى كنترل كيفيت
نين آزمايش مىشوند و با معيارهاى سادهاى مانند 3510 بعنوان هشدار و
براى رد نتايج» تفسير مىكردند. بعد از هر 5 روز نتايج 4510+
مجددا بازبينى و محدوده كنترل و ميانكين محاسبه م ىكردد.
صفحه 15:
حذف اعداد يرت
SS Aes ل ل ل
در محاسبه قرار میگیررد؛ باید معیاری بررای شناسائی و اصللاح خواندههای دور افتاده
تعيين نمود .روش ييشنهادى |
ا ا ا ا معیار" حاصله از بررسی قابلیت
تكرار يك سرى كار باشد(درمقايسه بال[51 1711:11112-11112آ در ينج روز اول) در
اين صورته ميبايد آسن <وانده ها حذف شده ويس از شناسائى و اصلاح منبع خطاء
انذازهكيررى مجددا انجام شود.
صفحه 16:
هرگاه بیش از 49 سسری های کار میباید حذف شوند و دلیل مشخصی
نیز بای این موارد شناسائی نشود ات
صفحه 17:
total number of days (generally 20)
run number within-day (1 or 2)
result for replicate 1, run j on day
result for replicate 2, run j on day
صفحه 18:
number of days (with two runs)
average result run 1, day i (average of the two replicates)
average result run 2, day i (average of the two replicates),
mumber of days
average of all results day i
average of all results.
صفحه 19:
A?
2
Sy=B’ -
where:
where:
= estimate of between-run standard deviation
صفحه 20:
صفحه 21:
تفسیر نتایج 202011107
square of the user’s estimated repeatability variance
square of the manufacturer’s claim of repeatability variance
the total number of runs (degrees of freedom for $1):
The calculated 3? should be compared with a statistical table of 7? values, using the upper 95% critical
vvalue with R degrees of fieedom (see Table 1). Ifthe calculated value is less than this table value, then
the estimate is not significantly different from the claimed value. and this part of the precision claim is
accepted,
صفحه 22:
102218 + 8+ MD)?
2ME? + MR? + MD
ME= S$, (mean square for within-run, or repeatability variance)
MR= 24° (mean square for runs)
MD= 4B? (mean square for days)
صفحه 23:
square of the user's estimate of w1tlun-laboratory (iwithmn-device) standard deviation
square of manufacturer’s claim of device standard deviation, or medically required standard
deviation
degrees of freedom for Sr.
Ifthe cae is sth the ented upper 95% val (om Table !( he piso perftmmance
ss astmned tobe acceptable
صفحه 24:
ا )- )0(
صفحه 25:
در اين دستورالعمل كه براى ارزيابى.دقت_توسط مصر فكننده
طراحی گردیده» روش سادهتری پیشنهاد شده است.
در این روش هم مانند روش قبل دورهآشنلیی با روش درنظرگرفته
شده است. تعداد.و غلظت مواد نيز بر اساس مواردى كه قبلا ذكر
ا ا ا 22062 ay i لمات
ا ا
صفحه 26:
۱
صفحه 27:
ا 0
مقدار © مزئئزب و زمر دارد.
Pe Ten ese 2036 ا ا
غلظت) در روز و براى 5 روز آزمايش مى شود.
اكر > © Pere . 3رپلیکه از هر نمونه (2
غلظت)در روزو براى 3 روز میشود.
صفحه 28:
۱
۱ ea epee Pm rrr eas)
5۲
اننا
501001 0
*2 : تعناد دفعاتةكرار آنمايثردر يكروز (3 يا4)
Pees eee alee) Gris
* 260 : میانگین کل نتایج
صفحه 29:
Total <4:
"شاخص واریانس 3 را بر اساس فرمول زیر محاسبه نمائید.
د Sel از راه زير حساب مىشود.
اک را 1۳۳
صفحه 30:
بررسی دقت بر اساس توصیه وستگارد
“در اين روش مىبايست دقت را در دو نوبت كوتاه مدت و بلند
مت ری رد ١
کوتاه مدت : 20 بار آزمایش هر ی
كارى :
صفحه 31:
CV % = SD*100/mean
صفحه 32:
صفحه 33:
مقایسه دقت دو روش ۱ ۱
*بسراى اينكه دقت و بعبارتى تكراريذيرى دو روش آزمايشكاهى با
بکدیگم مقایسه شود» میتوان از آزمون آماری ۳-165 استفاده
نمود. برای انجام این آزمون ابتدا یک نمونه با هر دو روش به
دفعات مورد آزمایش قرار میگیرد. pes eee ۱
نتایج محاسبه و در فرمول 1 قرار Soe میگردد.
صفحه 34:
"صرف نظر از نوع روش آزمایشگاهی. انحراف معیار بزرگتر در
صورت كسر و انحراف معيار كوجكتر در مخرج كسر قرار داده
ll casey ااا ا BOs oe eee lS
_
زوم “2
صفحه 35:
مواردى كه بايد توسط سازنده معرفى شوند:
*concentrations at which claim is made;
*repeatability standard deviation at every
0 16761
*repeatability percent coefficient of variation
*within-device (or within-laboratory) precision
standard deviation at every level tested;
*within-device or within-laboratory precision
percent coefficient of variation
‘actual number of days involved in the
experiment, and number of sites
صفحه 36:
*total number of observations
“number of instruments/devices used in the
evaluation, and how results were pooled;
* number of reagent lots; and
* number of calibration cycles and
calibration lots.
صفحه 37: